Distance Laser SensorDLS-A 15DLS-A 30DLS-AH 15DLS-AH 30Technical Reference ManualV1.06Please check www.dimetix.com for the latest version
TECHNICAL REFERENCE MANUAL 3 Installation3.1 MountingThree M4 threaded holes in the bottom of the DLS-A(H) make it easy to mount the device.Always obe
TECHNICAL REFERENCE MANUAL 3.2.3 Controlled modeRS232When using the RS232 interface only point-to-point communication is possible.❢ Never connect mult
TECHNICAL REFERENCE MANUAL 3.2.4 Automatic modeThe analog interface of the DLS-A(H) is isolated from the rest of the device. When using the analog int
TECHNICAL REFERENCE MANUAL 4 Technical data4.1 Measuring accuracyThe measuring accuracy corresponds to the ISO-recommendation ISO/R1938-1971 with a st
TECHNICAL REFERENCE MANUAL 4.2 SpecificationsTypical measuring accuracy for DLS-A 15 / DLS-AH 15 1)DLS-A 30 / DLS-AH 30 1)± 1.5 mm @ 2s± 3.0 mm @ 2sMa
TECHNICAL REFERENCE MANUAL 5 Electrical components5.1 ID switchThis switch is used to set the module ID and can be set from 0 to 9.5.2 Reset switchTo
TECHNICAL REFERENCE MANUAL 5.5 Connector5.5.1 D-Sub connectorPin Designator Description1 Rx RS232 receive line2 Tx RS232 send line3 T- RS422 send line
TECHNICAL REFERENCE MANUAL 6 Physical dimensionsAll dimension in mmDistance Laser Sensor Page 17/40150.0 38.5 88.5 Measuring reference = = 3 x M4 x 5
TECHNICAL REFERENCE MANUAL 7 Factory settings7.1 OperationMode: Controlled7.2 Communication parametersBaud: 19200Data bit: 7Parity: EvenStop bit: 17.3
TECHNICAL REFERENCE MANUAL 8 Command set8.1 General8.1.1 Command termination <trm>All commands for the DLS-A(H) are ASCII based and terminated &
Table of Contents1 Introduction...
TECHNICAL REFERENCE MANUAL 8.2.3 Temperature measurement (sNt)Triggers measurement of temperature.Command input:sNt<trm>N:Module number (0..9)Re
TECHNICAL REFERENCE MANUAL 8.2.6 Tracking with buffering – Start (sNf)Triggers continuous measurement of the distance with internal buffering in the m
TECHNICAL REFERENCE MANUAL 8.3 Configuration commands8.3.1 Set communication parameter (sNbr)Sets the communication parameters for the serial interfac
TECHNICAL REFERENCE MANUAL 8.3.3 Set analog output min level (sNvm)This command sets the minimum analog output current level (0 or 4 mA).Command input
TECHNICAL REFERENCE MANUAL 8.3.6 Set digital output levels (sNn)Sets the distance levels at which the digital outputs are switched ON and OFF with a h
TECHNICAL REFERENCE MANUAL 8.3.7 Save configuration parameters (sNs)This command saves all configuration parameters, which are set by the commands abo
TECHNICAL REFERENCE MANUAL 8.3.10 Get hardware version (sNN01N)Retrieves the hardware version of the DLS-A(H).Command input:sNN01N<trm>N:Module
TECHNICAL REFERENCE MANUAL 8.4 Command set for single module operation (Compatibility)The commands described in this chapter are compatible with comma
TECHNICAL REFERENCE MANUAL 8.4.5 Tracking (h)Triggers continuous measurement of the distance. The measurements are continued until the next command is
TECHNICAL REFERENCE MANUAL 8.4.9 Laser OFF (p)Switches laser OFF.Command input:p<trm>Return dataSuccessful:?<trm>Error: @Ezzz<trm>zz
TECHNICAL REFERENCE MANUAL 1 IntroductionThe DLS-A(H) is a powerful distance measuring instrument for integration into industrial applications. It all
TECHNICAL REFERENCE MANUAL 8.4.13 Date of manufacture (N03N)Outputs the date of manufacture at the interface.Command input:N03N<trm>Return dataS
TECHNICAL REFERENCE MANUAL 8.5 Error codesNo. Format Meaning203@E203wrong syntax in command, or prohibited command, or prohibited parameter incommand
TECHNICAL REFERENCE MANUAL 9 Accessories9.1 ViewfinderThe telescopic viewfinder can be used for easy alignment of the DLS-A(H) for long distances.Part
TECHNICAL REFERENCE MANUAL 9.4 CablesPart Number Description500200PC-Connection cable: DLS-A(H) to - 9 pin D-Sub for PC (RS232)- 2 wires for power sup
TECHNICAL REFERENCE MANUAL 10 Safety instructionsThe following directions should enable the person responsible for the DLS-A(H), and the user of the i
TECHNICAL REFERENCE MANUAL 10.2 Limits to use) See section ”Technical Data”Environment:Suitable for use in an atmosphere appropriate for permanent hum
TECHNICAL REFERENCE MANUAL 10.4 Hazards in useImportant hazards in useWARNING:The absence of instruction, or the inadequate imparting of instruction,
TECHNICAL REFERENCE MANUAL WARNING:Operate the equipment appropriately in accordance with the regulations in force. Always prevent access to theequipm
TECHNICAL REFERENCE MANUAL CAUTION:Looking into the laser beam may be hazardous to the eyes.Precautions:Do not look into the laser beam. Make sure the
TECHNICAL REFERENCE MANUAL 10.7 LabelingStandard applied EN60825-1:2001IEC60825-1:2001Emitted wavelength 620-690nmBeam divergence 0.16 x 0.6 mradPulse
TECHNICAL REFERENCE MANUAL 1.1 Product identificationThe product is identified by the serial label on the top of the enclosure:Version Typical Accurac
TECHNICAL REFERENCE MANUAL 10.8 MaintenanceThe DLS-A(H) is almost maintenance free. The only thing you have to do is cleaning the optical surfaces.CAU
TECHNICAL REFERENCE MANUAL 1.4 Measurement rangeThe DLS-A(H) is an optical instrument, whose operation is influenced by environmental conditions. Ther
TECHNICAL REFERENCE MANUAL 2 Device setupWe recommend performing the configuration steps in an office before mounting the device, especially if you ar
TECHNICAL REFERENCE MANUAL 2.2 Controlled modeIn controlled mode, each operation of a DLS-A(H) is triggered by a command sent from a host system over
TECHNICAL REFERENCE MANUAL 2.3 Automatic modeThe automatic mode is provided for host-less operation of the DLS-A(H). The analog and digitals outputs a
TECHNICAL REFERENCE MANUAL No. Action Comment Command5Save settings The changed configuration must besaved to make it permanent.Example for module 0:S