TROUBLESHOOTING
Issue Possible Cause Solution
DLS has no power
Poor cable connection Check connections between the DS8000
and the DLS100. If necessary, reseat the
cable in the socket.
Bent pin on 9-pin connector on
the DLS100 cable
Examine the 9-pin connector. Carefully
straighten any bent pins. If any pin is
missing, replace the cable (WPI #83016).
Yellow DLS100 On/Off switch is
set to off
Press the yellow button on the DLS100
No signal output (activity light off)
Green output switch on the
DS8000 is set to off
Set the output switch to on
DS8000 software “Run” check
box is not selected for the
appropriate channel
On the DS8000 screen, verify that
the Run and Output check boxes are
selected for the channel your DLS100 is
connected with
The appropriate channel check
box is not selected in the matrix
assignment screen on the
DS8000
Check the matrix assignment screen and
verify that the correct check boxes are
selected for the channel you are using.
See “DLS Control” on page 7.
Yellow DLS100 On/Off switch is
set to off
Press the yellow button on the DLS100
Non-compliance
alarm sounding
Open output circuit Check the integrity of all the connections
in your preparation
DC offset is greater than 20mV Adjust the DC offset. See “Adjusting the
DC Offset” on page 9.
NOTE: If you have a problem/issue with that falls outside the definitions of this
troubleshooting section, contact the WPI Technical Support team at 941.371.1003 or
technicalsupport@wpiinc.com.